Oxygen

PermeO2

Oxygen

FOR THIN FILM 

 

Oxygen permeability tester. PermeO2 is our model for measuring the oxygen transmission rate (OTR) through plastic films with a temperature and relative humidity controlled system.
PermeO2 is offered in two different versions: single cell, double cell.

Compliant with DIN 53380-3 norm
Compliant with ASTM F2622 norm

 

Tecnical specifications PermeO2

PERMEO2RS

Test range O2
0.01-15000 cm3·m-2·24h-1·bar-1
0.2-300000 cm3·m-2·24h-1·bar-1 (masked)
Test sample size
50 cm2 -  max tickness about 2,5 mm
Test temperature range
10-50 C ± 0.1 C
Relative humidity O2
0%, 5-95%, 100% ± 1.5%
Carrier flow (N2)
10-75 ml/min automatically controlled
Gas type N2
Pure gas (5.0 or 5.5 P.A.)
Gas type O2
Purity > 99.95%
Resdual humidity required < 0.5%
Gas pressure N2
1.5-2.0 bar
Gas pressure O2
1.5 bar
Gas connections (2)
Standard Ham-Let 1/8"
Software
LabView based
Desktop PC
Windows OS and LCD monitor
Power supply
100-240 VAC, 50/60 Hz (800 W max.)
Apparatus size (cm)
70 W/60 D/35 H whitout PC

 

 

PackPerm

Oxygen

FOR PACKAGING

 

Oxygen permeability tester. PackPerm is our model for measuring the oxygen transmission rate (OTR) through pouches, bags, little containers, PET bottles, natural or synthetic corks, screw and bottle tops, bag-in-box and other products for packaging application. Measures can be carried out both in air at room conditions and in 100% oxygen. PackPerm is offered in double cell version. At request it is possible to enclose the new technique of measure­ments patented by ExtraSolution that allows to reduce until 4 times the duration of a typical analysis.

Compliant with DIN 53380-3 norm
Compliant with ASTM F2622 norm

 

Tecnical specifications PackPerm

packpermRS

Test range O2
0.0001-25 cm3·pkg-1·24h-1·bar-1
Test sample size
PET bottles, closures, pouches, packages,....
Test temperature range
Room temperature
Relative humidity
In air: room humidity
In 100% O2: 0%
Carrier flow (N2)
10-75 ml/min automatically controlled
Gas type N2
Pure gas (5.0 or 5.5 P.A.)
Gas pressure N2
1.5-2.0 bar
Gas connections (2)
Standard Ham-Let 1/8"
Software
LabView based
Desktop PC
Windows OS and LCD monitor
Power supply
100-240 VAC, 50/60 Hz (800 W max.)
Apparatus size (cm)
53 W/58 D/30 H without PC
 